A World-Leading Supplier Of Medical Solutions
New generation of intelligent real-time output power feedback technology, excellent cutting performance,less damage to tissues, REM circuit detection safety system effectively avoids burns, protects patient safety.

– 4 Monopolar Cut: Pure Cut, Blend 1、2、3
– 2 Coag: Standard, Forced, Soft, Fine
– CQM Contact Quality Monitoring System
– Electrosurgical Pens and Foot Switch Control
– Recently Mode, Power and Other Parameters
– Quick Adjustment of Power and Volume
– Cut and Coag in an Intermittent Manner
Cone-cutting electrode can improved control effect during cone-cut biopsy

| Mode | Max Output Power(W) | Load Impedance(Ω) | Modulation Frequency(kHz) | Max Output Voltage(V) | Crest Factor | ||
| Monopolar | Cut | Pure Cut | 120 | 500 | --- | 1300 | 1.8 |
| Blend 1 | 120 | 500 | 20 | 1400 | 2.0 | ||
| Blend 2 | 120 | 500 | 20 | 1300 | 2.0 | ||
| Blend 3 | 100 | 500 | 20 | 1400 | 1.9 | ||
| Coag | Forced | 120 | 500 | 20 | 4800 | 6.2 | |
| Soft | 120 | 500 | 20 | 1000 | 2.0 | ||
| Bipolar | Standard | 100 | 100 | 20 | 700 | 1.9 | |
| Precise | 50 | 100 | 20 | 400 | 1.9 | ||